SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit

This product is intended for research purposes only.

The SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit utilizes isothermal amplification for use in the analysis of SARS-CoV-2, the novel coronavirus that causes COVID-19.

  • Colorimetric LAMP enables simple, visual detection (pink-to-yellow) of amplification of SARS-CoV-2 nucleic acid
  • Set up reactions quickly and easily, using a simple heat source and unique WarmStart® technology
  • Reduce risk of carryover contamination, with UDG and dUTP included in the master mix
  • Assay targets N and E regions of the SARS-CoV-2 genome, for optimized sensitivity and specificity
  • Bring confidence to your results using the provided controls
  • Learn more about LAMP and other isothermal amplification methods
  • Learn more about how NEB is supporting COVID-19 research
  • Need assistance designing LAMP primers? Use the NEB LAMP Primer Design Tool
 
 
Catalog # Concentration Size
E2019S Not Applicable 96 reactions
To find out how to order this product from your current location, click the button below:
International Customer & Order Support
Please enter a quantity for at least one size

Need a custom/large volume order? Contact us

Bulk packaging may also be available and requested for large recurring orders.
Learn More

  • Product Information


    The SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit utilizes Loop-Mediated Isothermal Amplification (LAMP) to detect SARS-CoV-2 nucleic acid. The kit is available for research use only and includes WarmStart Colorimetric LAMP 2X Master Mix with UDG and a primer mix targeting the N and E regions of the viral genome. Controls are provided to verify assay performance, and include an internal control primer set and a positive control template. Guanidine hydrochloride has been found to increase the speed and sensitivity of the RT-LAMP reaction and is also included.

     

    Figure 1: SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit Components

    Get results in just 30 minutes with NEB’s SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit.

    The SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit includes WarmStart Colorimetric LAMP 2X Master Mix with UDG and a primer mix targeting the N and E regions of the viral genome. Controls are provided to verify assay performance and include an internal control primer set and a positive control template. Guanidine hydrochloride has been found to increase the speed and sensitivity of the RT-LAMP reaction and is also included.



    Figure 2: SARS-CoV-2 Genome and SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit Gene Targets

     See all components in NEB’s SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit

    The SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit includes a dual primer mix that targets the N and E gene regions.



    WarmStart Colorimetric LAMP 2X Master Mix with UDG is an optimized formulation of Bst 2.0 WarmStart DNA Polymerase and WarmStart RTx in a special low-buffer reaction solution containing a visible pH indicator for rapid and easy detection of LAMP and RT-LAMP reactions. The inclusion of dUTP and UDG in the master mix reduces the possibility of carryover contamination between reactions.
    .
    This system provides a fast, clear visual detection of amplification based on the production of protons and the subsequent drop in pH that occurs from the extensive DNA polymerase activity in a LAMP reaction. The decrease in pH produces a change in solution color from pink to yellow. The total protocol time is 30 minutes and only requires a simple heating device that can reach 65°C.
    .
    A positive detection of the SARS-CoV-2 RNA sequence would be indicated by a yellow color (amplification occurred, protons released, pH-dependent color change from pink to yellow), while a negative result would be indicated by a pink color (no amplification occurred, no protons released, no color change).
    .
    Figure 3: SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit Workflow



    In the SARS-CoV-2 Colorimetric LAMP Assay Kit, three control reactions are run with each test sample. All reactions should be pink prior to incubation. Using the provided protocol, a Non-Template Control (NTC) reaction will contain all materials of the test sample (master mix, primers, etc) except for the test input nucleic acid itself and serves as a measure of reaction contamination and primer-based mis-amplification. The NTC sample should not amplify and should stay pink throughout the experiment. The Positive Control (PC) will contain master mix, a plasmid that contains the SARS-CoV-2 N-gene (GenBank: MN908947.3) and primers that will amplify this sequence. Amplification should be observed in the PC and the sample should become yellow after incubation. The Internal Control (IC) will contain master mix, test input nucleic acid, and LAMP primers for rActin, an endogenous housekeeping gene. If reagents are active and samples have been handled appropriately, the IC should become yellow after incubation.


     
    .
    Figure 4: Rapid, simple, color-based detection of SARS-CoV-2 RNA



    The SARS-CoV-2 Rapid Colorimetric LAMP Assay was carried out using the indicated controls and either positive sample (human total RNA + synthetic SARS-CoV-2 RNA) or negative sample (human total RNA alone). Valid results for Non-Template Control (NTC, pink), Positive Control (PC, yellow) and Internal Control (IC, yellow) are shown. In the positive test sample, isothermal amplification of SARS-CoV-2 RNA leads to a  color change from pink to yellow.



    .
    Figure 5: SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit Workflow



    The SARS-CoV-2 Rapid Colorimetric LAMP Assay was carried out using either positive samples (human total RNA + synthetic SARS-CoV-2 RNA at 5,000, 500 or 50 copies per reaction), negative samples (human total RNA alone) or no template (NTCs) as indicated. A. All positive samples gave a positive result with robust color change from pink to yellow, including detection in all samples at a low input of 50 copies per reaction (n = 20). B. No color change was observed for negative samples (n = 32). C. Summary of results from analogous testing conducted by 3 users in 9 experiments total. At 50 copies per reaction, >95% (277/288) of reactions were positive (yellow). For negative samples, >99% (358/360) of reactions gave no color change (pink) and 2/360 reactions gave a partial color change (orange, as indicated by asterisk for Users 2 and 3).


    LAMP has been used in novel workflows by scientists around the world to further COVID-19 research. To access application notes, selected preprints, and to learn more about NEB’s COVID-19 Researcher Spotlight series, click here.   

    Kit Components

    The following reagents are supplied with this product:

    Store at (°C) Concentration
    WarmStart® Colorimetric LAMP 2X Master Mix with UDG -20 2 X
    SARS-CoV-2 Positive Control (N gene) -20 12.5 X
    Internal Control LAMP Primer Mix -20 10 X
    SARS-CoV-2 LAMP Primer Mix (N/E) -20 10 X
    Nuclease-free Water -20
    Guanidine Hydrochloride -20 10 X
    Product Categories:
    Isothermal Amplification & Strand Displacement Products,
    PCR, qPCR & Amplification Technologies Products
    Applications:
    Isothermal Amplification,
    DNA Amplification, PCR & qPCR
    • Advantages and Features
    • Properties & Usage
    • Related Products
    • Product Notes
  • Protocols, Manuals & Usage
  • Tools & Resources
  • FAQs & Troubleshooting
  • Quality, Safety & Legal
    • Quality Control Assays
    • Specifications
    • Certificate of Analysis
    • Safety Data Sheets
    • Legal and Disclaimers

Loading Spinner